ISO/TR 18130:2016
p
ISO/TR 18130:2016
61530

Status : Published

en
Format Language
std 1 63 PDF
std 2 63 Paper
  • CHF63
Convert Swiss francs (CHF) to your currency

Abstract

ISO/TR 18130:2016 provides guidelines for a method to determine the extent of screw loosening of the metallic implant body/implant abutment joint of endosseous dental implants, such as two-part implants or multi-part implants under cyclic torsional loading. This test is most appropriate for evaluating new types of joints fixed using screw(s) and metallic connecting parts. This Technical Report provides a protocol for torsional cyclic torque on an implant body/implant abutment joint, but its intended use is for prefabricated implant bodies, implant abutments and, if appropriate, implant connecting parts that are made of metallic materials.

It is not applicable to ensure the in vivo performance of endosseous dental implants and is not derived from observations of clinical failures.

NOTE This Technical Report is not intended for use with temporary abutments.

Read sample 

Preview this standard in our Online Browsing Platform (OBP)

General information

  •  : Published
     : 2016-05
    : International Standard published [60.60]
  •  : 1
     : 7
  • ISO/TC 106/SC 8
    11.060.15 
  • RSS updates

Got a question?

Check out our FAQs

Customer care
+41 22 749 08 88

Opening hours:
Monday to Friday - 09:00-12:00, 14:00-17:00 (UTC+1)